Medical-Grade Thermometer

ACTAtek Temsen

Medical-grade contactless infrared body thermometer for high-precision, fast, and safe temperature screening at entry points.

ACTAtek Temsen Temperature Sensor Product Shot

Overview

ACTAtek Temsen is a medical-grade infrared human body thermometer designed for accurate, safe, fast, and contactless temperature measurement. Features a high-precision proximity sensor ensuring objects are within measurable range. Ideal for environments requiring precise and reliable temperature monitoring including offices, hospitals, schools, and public venues.

Technical Specifications

ParameterValue
Power Supply12VDC (from ACTAtek DC output)
InterfaceRS232 or RS485
Measuring Range35.0°C – 42.8°C
Abnormal Alert Threshold37.5°C – 38°C
LED IndicationRed (alert) / Green (normal)
Accuracy±0.2°C
Speed600 ms per reading
Dimensions8.0 cm (W) × 2.7 cm (H) × 3.5 cm (D)

AMS Healthcare Integration

Temsen integrates directly with the AMS Healthcare module to provide:

Automated health screening at every entry point with zero user contact
Dynamic access control: deny entry if temperature exceeds threshold
Contact tracing reports linked to individual biometric identity
Crowd capacity monitoring with health-based restrictions
Risk group segregation — restrict high-risk individuals to specific areas
Real-time alerts to security / HR when abnormal temperature detected
Temsen connects to the ACTAtek terminal via the RS232 or RS485 port, powered directly from the terminal's DC OUT port (12VDC). No additional power supply is required.

Installation Notes

  1. Connect Temsen to the ACTAtek terminal's RS232 or RS485 port.
  2. Power the Temsen from ACTAtek's DC OUT (12V, 1A).
  3. Position the Temsen at a comfortable height (approximately 130–160 cm from floor for adult forehead level).
  4. Avoid placement near heat sources, air vents, or direct sunlight to prevent false readings.
  5. Configure abnormal temperature threshold (default 37.5°C) in AMS Healthcare settings.
  6. Test with a calibrated reference thermometer before deployment.